Operation

HandyLab 100

38

ba77062d01

03/2014

Note
For single-point calibration, the instrument uses the Nernst slope
(-59.2 mV/pH at 25 °C) and determines the zero point of the electrode.

Continuing with two-

point calibration

6

Start the measurement with <ENTER>.
The measured value is checked for stability (stability control).
The [AR] status indicator is displayed. The measured parame-
ter flashes.

7

Wait for the end of the measurement with stability control or
accept the calibration value with <ENTER>.
The calibration display for the next buffer appears (voltage dis-
play).

8

If necessary, finish the calibration procedure as a single-point
calibration with <M>.
The calibration record is displayed.

pH

-180.0

24.8 °C

AutoCal TEC

10.011

15.03.2014

010:00

AR

Buffer

mV

9

Thoroughly rinse the electrode with deionized water.

10

Immerse the electrode in buffer solution 2.

11

When measuring without temperature sensor:
Measure the temperature of the buffer manually and enter it
with <

><>.

12

Start the measurement with <ENTER>.
The measured value is checked for stability (stability control).
The [AR] status indicator is displayed. The measured parame-
ter flashes.