Operation
HandyLab 100
40
ba77062d01
03/2014
Note
Calibration is automatically completed after the last buffer of a buffer
set has been measured. Then the calibration record is displayed.
The calibration line is determined by linear regression.
19
Wait for the measurement with stability control to be completed
or terminate the stability control and take over the calibration
value with <ENTER>.
The calibration display for the next buffer appears (voltage dis-
play).
20
If necessary, use <M> to finish calibration or
switch to calibration with the next buffer with <ENTER>.