Overview

handylab pH/LF 12

10

ba75370e03

11/2009

Measuring ranges

of conductivity

Precision of

conductivity (± 1

digit)

Cell constant,

calibrating

Reference

temperature of

conductivity

Measuring range

Resolution

ӕ

[µS/cm]

0 ... 1999

1

ӕ

[mS/cm]

0.00 ... 19.99
0.0 ... 199.9
0 ... 500

0.01
0.1
1

SAL

0.0 ... 70.0 accord-
ing to the IOT table

0.1

T [°C]

− 5.0 ... + 105.0

0.1

ӕ

Nonlinear compensation

:

Precision Sample

temperature

± 0.5 %

0 °C ... 35 °C
according to EN 27 888;

± 0.5 %

35 °C ... 50 °C
extended nLF function

Linear compensation
Accuracy Sample

temperature

± 0.5 %

10 °C ... 75 °C

(The accuracy percentage always refers to
the measured value.)

SAL

Range 0.0 ... 42.0
Accuracy Sample

temperature

± 0.1

5 °C ... 25 °C

± 0.2

25 °C ... 30 °C

T [°C]

± 0.1

C [cm

-1

]

0.450 ... 0.500

Tref

Can be set to 20 °C or 25 °C