Xylem handylab pH LF 12 User Manual

Page 74

Advertising
background image

Lists

handylab pH/LF 12

74

ba75370e03

11/2009

Glossary

Adjusting

To manipulate a measuring system so that the relevant value (e. g. the
displayed value) differs as little as possible from the correct value or
a value that is regarded as correct, or that the difference remains with-
in the tolerance.

Asymmetry

Designation for the offset potential of a pH electrode. It is the measur-
able potential of a symmetrical electrode, the membrane of which is
immersed in a solution with the pH of the nominal electrode zero point
(electrodes: pH = 7).

AutoRange

Name of the automatic selection of the measuring range.

AutoRead

Name for a function to check the stability of the measured value.

Calibration

Comparing the value from a measuring system (e. g. the displayed
value) to the correct value or a value that is regarded as correct. Of-
ten, this expression is also used when the measuring system is adjust-
ed at the same time (see adjusting).

Cell constant, k

Characteristic quantity of a conductivity measuring cell, depending on
the geometry.

Conductivity

Short form of the expression, specific electrical conductivity. It is a
measured value of the ability of a substance to conduct an electric cur-
rent. In water analysis, the electrical conductivity is a dimension for the
ionized substances in a solution.

Conductometry

Name of the conductivity measuring technique.

Diaphragm

The junction is a porous body in the housing wall of reference elec-
trodes or electrolyte bridges. It forms the electrical contact between
two solutions and makes electrolyte exchange more difficult. The ex-
pression, junction, is also used for ground or junction-less transitions.

Electrode zero point

The zero point of a pH electrode is the pH value at which the electro-
motive force of the pH electrode at a specified temperature is zero.
Normally, this is at 25 °C.

Electromotive force of

an electrode

The electromotive force U of the electrode is the measurable electro-
motive force of an electrode in a solution. It equals the sum of all the
galvanic voltages of the electrode. Its dependency on the pH results
in the electrode function which is characterized by the parameters,
slope and zero point.

Measured parameter

The measured parameter is the physical dimension determined by
measuring, e. g. pH, conductivity or D. O. concentration.

Measured value

The measured value is the special value of a measured parameter to
be determined. It is given as a combination of the numerical value and
unit (e. g. 3 m; 0.5 s; 5.2 A; 373.15 K).

Advertising