3 concal, Concal, S y a – Xylem handylab pH LF 12 User Manual

Page 32

Advertising
background image

Operation

handylab pH/LF 12

32

ba75370e03

11/2009

4.3.3

ConCal

Two-point

calibration

For this procedure, use two buffer solutions:

z pH 7.0 ± 0.5
z any other buffer solution

Note
Skip the steps 2 and 9 if you use a pH electrode with a temperature probe or
the temperature probe of a conductivity measuring cell.

Note
At this point, the ConCal calibration can be broken off with <M>. This corre-
sponds to a single-point calibration. When doing so, the instrument uses
the Nernst slope (-59,2 mV/pH at 25 °C) and the fixed asymmetry of the elec-
trode.

1

Press the <CAL> key repeatedly until the ASY display and the
Cal function display appears. The sensor symbol displays the
evaluation of the last calibration (or no sensor symbol in the de-
livery state or after the measurement parameter has been re-
set).

2

If required, enter the temperature of the first buffer solution with
<

> <> while keeping the <RUN/ENTER> key depressed.

3

Immerse the pH electrode in the first buffer solution (pH 7.0 ±
0.5 in two-point calibration).

4

Press the <RUN/ENTER> key.
The measured pH value appears on the display.

5

Set the measured value to the nominal pH value of the buffer
solution (at the current temperature) with <

> <>.

6

When the measured value is stable, press the <RUN/ENTER>
key. The value of the asymmetry appears.

7

Press the <RUN/ENTER> key. SLO appears.

/

K

cm

Sal

1

S

O

U

8

°

F

%

1/cm

mV/pH

S/

m

cm

M

%

mg/l

mbar

Lin

Tref25

nLF

Time

Baud

Day.Month No.

Ident

Year

LoBat

AutoCal DIN

AutoCal TEC

RCL

Auto

Store

ARng

AR

TDS

Sal

Tref20

8

4

2

S Y

A

pH

Cal

°

C

TP