handylab pH/LF 12

Operation

19

ba75370e03

11/2009

Temperature

measurement in

pH measurements

You can perform pH measurements with or without a temperature probe as
well as with the temperature probe of a conductivity measuring cell. The mea-
suring instrument recognizes which sensors are connected and automatically
switches to the correct mode for the temperature measurement.

The following cases are distinguishable.

* If you do not wish that, you can:

– either disconnect the 2nd sensor and use the manual temperature input

or

– use an electrode with a temperature probe.

If a temperature probe is connected, it is indicated on the display by TP.

Note
When calibrating without a temperature probe (no TP display indicator dis-
played), enter the current temperature of the respective buffer solution man-
ually using the <

> <> keys while keeping the

<RUN/ENTER> key depressed.

Temperature probe

Display

Mode

pH

Cond

yes

-

TP

Automatic with
pH temperature probe

yes

yes

TP

-

-

Manual

-

yes

TP
flashes

The temperature value of the
2nd sensor (Cond) in the
same test sample is used for
the pH measurement*