9 reset, Reset – Xylem handylab pH LF 12 User Manual
Page 62
Operation
handylab pH/LF 12
62
ba75370e03
11/2009
4.9
Reset
You can reset (initialize) the measurement parameters and the configuration
parameters separately from one another.
Measurement
parameters
The following measuring parameters are reset to the delivery status:
pH
Conductivity
Note
The calibration data gets lost when the measuring parameters are reset.
Recalibrate after performing a reset.
Configuration
parameters
The following configuration parameters (InI) are reset to the delivery status:
Measuring mode
pH
Slope -59.16 mV/pH
Asymmetry
0 mV
Manual temperature input
25 °C
Measuring mode
ӕ
Cell constant
0.475 cm
-1
(calibrated)
0.475 cm
-1
(set up)
Temperature compensation
nLF
Reference temperature
Temperature coefficient of the lin-
ear temperature compensation
2.000 %/K
Baud rate
4800
Display during the
pH calibration
Buffer nominal value
Interval 1
(automatic save)
OFF
Interval 2
(for data transmission)
OFF