9 reset, Reset – Xylem handylab pH LF 12 User Manual

Page 62

Advertising
background image

Operation

handylab pH/LF 12

62

ba75370e03

11/2009

4.9

Reset

You can reset (initialize) the measurement parameters and the configuration
parameters separately from one another.

Measurement

parameters

The following measuring parameters are reset to the delivery status:

pH

Conductivity

Note
The calibration data gets lost when the measuring parameters are reset.
Recalibrate after performing a reset.

Configuration

parameters

The following configuration parameters (InI) are reset to the delivery status:

Measuring mode

pH

Slope -59.16 mV/pH

Asymmetry

0 mV

Manual temperature input

25 °C

Measuring mode

ӕ

Cell constant

0.475 cm

-1

(calibrated)

0.475 cm

-1

(set up)

Temperature compensation

nLF

Reference temperature

Temperature coefficient of the lin-
ear temperature compensation

2.000 %/K

Baud rate

4800

Display during the
pH calibration

Buffer nominal value

Interval 1
(automatic save)

OFF

Interval 2
(for data transmission)

OFF