2 conductivity system messages, 3 general errors, Conductivity system messages – Xylem handylab pH LF 12 User Manual

Page 69: General errors

Advertising
background image

handylab pH/LF 12

What to do if...

69

ba75370e03

11/2009

6.2

Conductivity system messages

Error message

OFL

Error message

E3

6.3

General errors

Display LoBat

Instrument does

not react

to keystroke

Display

to

Probe symbol

flashes

Cause

Remedy

The measured value lies outside
the measuring range

– Measuring cell not connected

– Connect measuring cell

– Cable broken

– Replace measuring cell

Cause

Remedy

– Measuring cell contaminated

– Clean cell and replace it if

necessary

– Unsuitable calibration

solution

– Check calibration solutions

Cause

Remedy

– Batteries almost empty

– Exchange the batteries (see

section 5.1 M

AINTENANCE

)

Cause

Remedy

– Operating condition undefined

or EMC load unallowed

– Processor reset:

Switch the instrument on
while pressing the <CAL> key

Cause

Remedy

– Time-out of the interface

– Check the instrument that is

connected

Cause

Remedy

– Calibration interval expired

– Recalibrate the measuring

system