handylab pH/LF 12

Operation

59

ba75370e03

11/2009

4.8

Configuration

You can adapt the measuring instrument to your individual requirements. To
do this, the following parameters can be changed (the status on delivery is
marked in bold):

Note
You can leave the configuration menu at any time with <M>. The parameters
that have already been changed are stored.

Baudrate

Baud rate

1200, 2400, 4800, 9600

Display during the
pH calibration

Buffer nominal value, or current
electrode voltage

Intervals
Calibration (Int 3)
in days (d)

– pH: 1 ... 7 ... 999 d

Int 3

: 1 ... 180 ... 999 d

Int 5

Reference temperature

– 25 °C (TREF25) t25

– 20 °C (TREF20) t20

Date/time

Any

1

Switch off the measuring instrument.

2

Press the <M> key and hold it down.

3

Press the <ON/OFF> key.
The display test appears briefly on the display. The measuring
instrument then switches automatically to the setting of the
baud rate.

4

Select the required Baud rate with <

> <>.

5

Confirm with <RUN/ENTER>. CAL dISP appears on the dis-
play.

Day.Month No.

Time

1

2

M

m

%

mg/l

cm

S/

cm

S

mbar

mV/pH

t

n

Store

Auto

LoBat

Tref25

nLF

/

K

%

RCL

Sal

°

AR

TP

C

cm

1/

°

F

ARng

Cal

Lin

Tref20

Sal

TDS

O

UpH

AutoCal DIN

AutoCal TEC

Ident

Year

0

0

8

4

Baud