handylab pH/LF 12

36

ba75370e03

11/2009

Checking the cell

constant

1

Press the <RCL> key repeatedly until CAL disp appears on the
display.

2

Press the <RUN/ENTER> repeatedly until the last calibrated
cell constant is displayed, e. g. 0.472 cm

-1

.

3

To return to the measuring mode: Press the <M> key when the
correct cell constant is displayed.

4

If you want to recalibrate the cell constant, proceed according
to section 4.4.5 D

ETERMINING

THE

CELL

CONSTANT

(C

ALIBRATION

IN

THE

CONTROL STANDARD

).