Operation

handylab pH/LF 12

46

ba75370e03

11/2009

4.5

Calibration intervals (Int 3, Int 5)

For each measured variable, a time interval is stored. When it has expired,
you will be reminded to calibrate. After a calibration interval has expired, the
probe symbol of the relevant measured variable flashes. It is still possible to
measure. By calibrating the relevant probe, the function is reset and the in-
terval starts anew.

The following calibration intervals are set in the factory:

Setting the

calibration interval

You can change each of these intervals (1 ... 999 days):

Measured
parameter

Designation

Default setting

pH/ORP

Int 3

7 days (d 7)

Conductivity

Int 5

180 days (d 180)

1

Switch off the measuring instrument.

2

Press the <M> key and hold it down.

3

Press the <ON/OFF> key.
The display test appears briefly on the display. After this, the
measuring instrument automatically switches over to configu-
ration.

4

Press the <RUN/ENTER> key repeatedly until Int 3 or Int 5 and
the required measured variable pH or

appears. Example:

5

Set the required interval (in days) until the next calibration with
<

> <>.

6

Confirm with <RUN/ENTER>.

7

Switch to the measuring mode with <M>.

O

8

1

°

C

TP

ARng

mg/l

LoBat

Baud

Day.Month No.

Ident

Year

Sal

%

mbar

S

RCL

AR

Store

Auto

°

F

1/cm

/

K

%

TDS

Sal

mV/pH

S/

m

cm

cm

M

Lin

Tref25 Tref20

nLF

AutoCal DIN

AutoCal TEC

7

t 3

n

1

d

Time

Cal

pH

Advertising