Operation

handylab pH/LF 12

56

ba75370e03

11/2009

4.6.4

Clearing the memory

With this function, you can delete stored data records. 500 storage locations
will then be available again.

Note
The Clear memory function only appears when there are data records stored
in the memory. Otherwise, the measuring instrument automatically switches
to the last active measuring mode.

Proceed as follows to clear all data records:

Note
The calibration data remain stored and can be called up.

1

Switch off the measuring instrument.

2

Press the <STO> key and hold it down.

3

Press the <ON/OFF> key.
The display test appears briefly on the display.

4

Confirm the clearing process with <RUN/ENTER>.
Pressing any other key prevents the clearing, the data records
will remain stored.