2 measuring, 1 general information, Measuring 4.2.1 – Xylem Cond 1970i User Manual

Page 16: General information

Advertising
background image

Operation

Cond 1970i

16

ba75692e01

01/2007

4.2

Measuring

4.2.1

General information

You can measure the following variables:

z Conductivity
z Salinity
z Total dissolved solids (TDS)

The measuring instrument is supplied with the following functions:

z AutoRange (automatic switchover of the measurement range). If a

measuring range is exceeded, AutoRange causes the measuring in-
strument to change automatically to the next higher measuring
range and back again. Therefore, the instrument always measures
in the measuring range with the highest possible resolution. The
function can be switched off.

z The AutoRead function (drift control) for checking the stability of the

measurement signal. This ensures the reproducibility of the measur-
ing signal. For details of how to switch the AutoRead function on/off,
see page 20.

Preparatory activities

Perform the following preparatory activities when you want to measure:

Warning
When connecting an earthed PC/printer, measurements cannot be per-
formed in earthed media as incorrect values would result. The RS232
interface is not galvanically isolated.

Temperature

compensation and

reference temperature

Tref

The calculation of the temperature compensation is based on the pre-
set reference temperature, 20 °C or 25 °C. For more detailed informa-
tion concerning the temperature compensation, see section 4.4 on
page 28.

1

Connect a conductivity measuring cell to the measuring instru-
ment.

2

Calibrate or check the measuring instrument with the measur-
ing cell. How to calibrate is described in section 4.3.

3

Select the measuring mode with <M>.

Advertising