Cond 1970i

Operation

43

ba75692e01

01/2007

4.6.2

Recorder (analog output)

You can transmit data to a recorder via the analog output. Connect the
analog output to the recorder via the AK323 interface cable.
The data output automatically switches to Recorder output.

Socket assignment

Note
The analog output is activated automatically in the cable by connecting
2 and 3.

The output on the analog output corresponds to the value shown on the
display.

Signal range

The signal range of the analog output depends on the measured vari-
able and the measuring range:

Conductivity

Salinity

1 free
2 Plug coding
3 Ground
4 Analog output

1

2

3

4

RS 232

REC

(internal resistance < 5 Ohm)

Measuring range

Voltage

Resolution

0.000 ... 1.999 µS/cm

0 ... 1999 mV

1 mV

0.00 ... 19.99 µS/cm

0 ... 1999 mV

1 mV

0.0 ... 199.9 µS/cm

0 ... 1999 mV

1 mV

0 ... 1999 µS/cm

0 ... 1999 mV

1 mV

0.00 ... 19.99 mS/cm

0 ... 1999 mV

1 mV

0.0 ... 199.9 mS/cm

0 ... 1999 mV

1 mV

0 ... 500 mS/cm

0 ... 500 mV

1 mV

Measuring range

Voltage

Resolution

0 ... 70.0

0 ... 700 mV

1 mV