5 autoread ar (drift control) and hold function, Autoread ar (drift control) and hold function – Xylem Cond 1970i User Manual

Page 20

Advertising
background image

Operation

Cond 1970i

20

ba75692e01

01/2007

4.2.5

AutoRead AR (drift control) and hold function

The AutoRead function (drift control) checks the stability of the mea-
surement signal. The stability has a considerable impact on the repro-
ducibility of the measured values. With the aid of the hold function the
measured value display is frozen.

Use the AutoRead function and hold function like this:

Note
The current AutoRead measurement can be terminated at any time
(accepting the current value) by pressing <RUN/ENTER>.

1

Select the required measuring mode with <M>.

2

Activate the AutoRead function with <AR>.
The current measured value is frozen (hold function).

3

Start AutoRead with <RUN/ENTER>.
AR flashes until a stable measured value is reached.

4

If necessary, start the next AutoRead measurement with
<RUN/ENTER>.

5

To terminate AutoRead: Press the <AR> key.