Printing the

calibration protocol

The calibration protocol contains the calibration data of the current cal-
ibration. You can transmit the calibration protocol to a printer via the se-
rial interface (see page 39).

Note
You can automatically print a calibration protocol after the calibration.
To do so, connect a printer to the interface according to section 4.6.3
before calibrating. After a valid calibration, the record is printed.

Sample printout:

Calibration evaluation

After the calibration, the measuring instrument automatically evaluates
the current status of the calibration. The evaluation appears on the dis-
play.

CALIBRATION PROTOCOL
14.04.02 11:37
Device No.: 99990000
Calibration Conductivity
Cal time: 14.04.02 / 11:37
Cal interval: 180d
Cal Std.: 0.01 mol/l KCL
40.0 °C
Conduct./Tref25: 1413µS/cm
Cell Const : 0.478 1/cm
Probe : +++

Display

Cell constant [cm

-1

]

0.450 ... 0.500 cm

-1

0.800 ... 1.200 cm

-1

E3

Eliminate the error according to
chapter 6 W

HAT

TO

DO

IF

...

outside the ranges
0.450 ... 0.500 cm

-1

or
0.800 ... 1.200 cm

-1