C t 1 1, E l l c – Xylem Cond 1970i User Manual
Page 27

Cond 1970i
Operation
27
ba75692e01
01/2007
Range
0.250 ... 2.500 cm
-1
1
Press the <CAL> key repeatedly until CELL appears on the
display.
2
Press the <RUN/ENTER> key.
3
Press the <CAL>
repeatedly until until a cell constant in the
range 0.250 ... 2.500 cm
-1
appears.
4
Set the cell constant to be used with <
▲> <▼>,
e.g. 0.614 cm
-1
.
5
To return to the measuring mode: Press the <M> key.
From now on, the new cell constant will be used.