6 what to do if, What to do if, Chapter 6 w – Xylem Cond 1970i User Manual

Page 53: 6what to do if

Advertising
background image

Cond 1970i

What to do if...

53

ba75692e01

01/2007

6

What to do if...

Display LoBat

Instrument does not

react to keystroke

Error message

OFL

OFL

OFL

OFL

Error message

E3

E3

E3

E3

Display

to

to

to

to

Cause

Remedy

– Battery almost empty

– Charge the battery (see

section 3.2)

Cause

Remedy

– Operating condition undefined

or EMC load unallowed

– Processor reset:

Press the <RCL> and
<ON/OFF> keys at the same
time and release them again.
The software version is
displayed.

Cause

Remedy

The measured value lies outside
the measuring range

– Measuring cell not connected

– Connect measuring cell

– Cable broken

– Replace measuring cell

Cause

Remedy

– Measuring cell contaminated

– Clean cell and

replace it if necessary

– Unsuitable calibration

solution

– Check calibration solutions

Cause

Remedy

– Time-out of the interface

– Check the instrument

connected