2 setting the cell constant manually, Setting the cell constant manually, C t 1 1 – Xylem Cond 1970i User Manual
Page 25: E l l c

Cond 1970i
Operation
25
ba75692e01
01/2007
4.3.2
Setting the cell constant manually
Note
The cell constant to be set must either be taken from the operating
manual of the measuring cell or is printed on the measuring cell.
You can set the cell constant manually as follows:
Setting the fixed cell
constant
0.010 cm
-1
You can set the fixed value 0.010 cm
-1
for the cell constant as follows:
1
Press the <CAL> key repeatedly until CELL appears on the
display.
2
Press the <RUN/ENTER> key.
3
Press the <CAL>
key repeatedly until the cell constant
0.010 cm
-1
appears on the display.
4
To return to the measuring mode: Press the <M> key.
From now on, the cell constant 0.010 cm
-1
will be used.