1-6

Section 1 - Introduction

Installation and Operation Manual

X-TMF-SLAMf-MFC-eng

Part Number: 541B032AAG

April, 2013

Brooks

®

Digital MFC's and MFM's

Figure 1-1 Open Collector Alarm Output

Figure 1-2 General Wiring

Valve Override Signal Specifications (Analog I/O Versions only)

Input Impedence: 800 k

Ω

Absolute Max. Input: -25 Vdc > VOR > 25 Vdc
(without damage)

5 Volt Reference Signal (Analog I/O versions only)

A 5 Vdc reference output is provided to the customer for use in generating
a setpoint and/or Valve Override signal. The current drive of this output is
very limited and must be used with care.

Min. Load Resistance: 2 k

Ω (2.5 mA maximum)

Accuracy: +1.0%