Section 4 Maintenance &
Troubleshooting

Installation and Operation Manual

X-TMF-SLAMf-MFC-eng

Part Number: 541B032AAG

April, 2013

Brooks

®

Digital MFC's and MFM's

Remove the sensor connector from the PC Board for this procedure.

OHMMETER CONNECTION

RESULT IF ELECTRICALLY FUNCTIONAL

Open circuit on ohmmeter. If either heater (1) or

Pin 1 or 4 to meter body

sensor common (4) are shorted, an ohmmeter
reading will be obtained.

Pin 4 to Pin 2

Nominal 1100 ohms reading, depending on

Pin 4 to Pin 3

Pin 5 to Pin 1

Nominal 1000 ohm reading.

Pin 6 to Pin 7

Nominal 580 ohm reading.

temperature and ohmmeter current.

SENSOR

SCHEMATIC

PIN

NO.

FUNCTION

1

Heater

Upstream

2

Temperature

Sensor (Su)

Downstream

3

Temperature

Sensor (Sd)

4

Sensor Common

5

Heater Common

6

Thermistor

7

Thermistor

Table 4-1 Sensor Troubleshooting

Flex Circuit Wire Numbers

4-1-3 Calibration Procedure

The calibration of Brooks Digital Series Mass Flow devices is not described
in this manual. Such calibration requires accurate and traceable calibration
equipment in addition to digital communications.
If your device needs calibration Brooks Instrument can provide this service
at one of its service locations. Visit www.BrooksInstrument.com to locate
the service location nearest to you. However, if traceable calibration
equipment is available at your facility, Brooks service Suite Calibration
software, along with training, is available for purchase.