Section 4 Maintenance &

Troubleshooting

Installation and Operation Manual
X-TMF-SLAMf-MFC-eng
Part Number: 541B032AAG
April, 2013

Brooks

®

Digital MFC's and MFM's

Table 4-2 Troubleshooting

Trouble

Possible cause

Check/Corrective Action

Output stays at zero

Clogged Sensor

Clean sensor. Refer to cleaning procedures

(regardless of Setpoint) and

(Section 4-1-2).

there is flow through the
meter/controller

Defective electronic board

Contact Brooks Instrument

Flow cannot be achieved

Clogged Control Valve

Clean the control valve (Section 4-1-2) or

regardless of Setpoint.

return the device to the factory

(applicable to MFC)

Valve override input is grounded

Check the valve override input (Pin 12)

Defective electronic board

Contact Brooks Instrument.

Output signal stays at approx.

Valve leaks or is stuck open

Clean and/or adjust control valve

5.5 Vdc or 22 mA

(applicable to MFC)

(Section 4-1-2).

(regardless of Setpoint) and
there is flow through the

+15 V applied to the valve

Check the valve override terminal (Pin 12)

meter/controller

override input
(applicable to MFC)

Defective PC board

Contact Brooks Instrument

Output signal follows Setpoint

Control valve leaks or is stuck

Clean the control valve or return the device

at higher Setpoints but will not

open.

to the factory (Section 4-1-2).

go below 2%
Output signal follows Setpoint

Insufficient inlet pressure or

Adjust pressures, inspect in-line filters and

at lower Setpoints, but does

pressure drop

clean/replace as necessary.

not reach full scale

Partially clogged sensor

Clean sensor, see cleaning procedures
(Section 4-1-2).

Partially clogged valve

Clean the control valve (Section 4-1-2) or

(applicable to MFC)

return the device to the factory, see
cleaning procedures

Valve out of adjustment

Contact Brooks Instrument

(applicable to MFC)

Valve guide spring failure

Contact Brooks Instrument

(applicable to MFC)

Instrument grossly out of

Partially clogged sensor

Clean sensor, see cleaning procedures

calibration. Flow is higher than

(Section 4-1-2).

desired.
Instrument grossly out of

Partially clogged restrictor

Replace or clean restrictor

calibration. Flow is lower than
desired.
Controller oscillates

Pressure drop or inlet pressure

Adjust pressures to original specifications

(applicable to MFC)

deviates from calibrated values

Valve out of adjustment

Contact Brooks Instrument

Unstable inlet pressure

Check external pressure regulator

Defective PC board

Contact Brooks Instrument