ST100 Series Flow Meter

INSTALLATION

Fluid Components International LLC

7

Flange Mount

The flange mount flow element in shown in Fig 3. Attach the process mating flange with care. The correct orientation of the flow element must be

maintained to ensure the calibrated accuracy.
• Verify the process media flow matches the flow direction arrow on the flow element.
• Apply appropriate gasket and or sealant to flange mount as required.
• Mate flow element flange to process flange keeping flat oriented properly.
• Secure flanges with appropriate mounting hardware.

NPT Pipe Thread Mount

The pipe thread configuration is shown in Fig 4. Apply sealant compatible with the process media to male threads. Carefully insert into process

mounting coupling. Tighten the flow element until snug and continue until flat and flow direction arrow are aligned with process flow.