Technical specification – Fluid Components International ST100 Series User Manual

Page 7

Advertising
background image

ST100 Series Flow Meter

GENERAL

Fluid Components International LLC

3

Instrument

„

Measuring Capability

ST1XX Models: Flow rate, total flow and temperature

STP1XX Models: Flow rate, total flow, temperature and pressure

„

Basic Style

ST100:

Insertion, single-point

ST100L:

In-line (spool piece), single-point

ST102:

Insertion, dual-element system

ST110:

Insertion, single-point with VeriCal™ capability

ST112:

Insertion, dual-element system with VeriCal capability

STP100: Insertion, single-point with pressure measurement

STP102: Insertion, dual-element system with pressure measure-

ment

STP110: Insertion, single-point with pressure measurement and

VeriCal capability

STP112: Insertion, dual-element system with pressure measure-

ment and VeriCal capability

„

Flow Measurement Range

Insertion Style: 0.25 SFPS to 1000 SFPS [0,07 NMPS to 305 NMPS]

ST100L In-line: 0.0062 SCFM to 1850 SCFM

[0.01 Nm3/h to 3,140 Nm3/h]

– Air at standard conditions; 70 °F and 14.7 psia [0 °C and 1,01325

bar (a)]

„

Temperature Measurement Range

Up to 850 °F [454 °C] commensurate with element; see Operating

Temperature in Flow Element specification

„

Pressure Measurement Range (STP Models)

Available Ranges:

0 psig to 50 psig [0 bar (g) to 3,4 bar (g)]

0 psig to 160 psig [0 bar (g) to 11 bar (g)]

0 psig to 500 psig [0 bar (g) to 34 bar (g)

0 psig to 1000 psig 0 bar (g) to 70 bar (g)]

„

Media:

All gases that are compatible with the flow element material

„

Accuracy:

Flow:

Gas Specific Calibration: ± 0.75% reading, ± 0.5% full scale

SpectraCal Gas Equivalency: Typically ± 4% reading, ± 0.5% full

scale;

gas conditions specific to application will determine accuracy; utilize

FCI’s online tool, AVAL, to evaluate your application and provide

expected accuracy

Temperature: ± 2 °F [± 2 °C] (display only, flow rate must be greater

than 5 AFPS [1,5 m/sec])

Pressure (STP Models): ± 0.25% full scale pressure range

„

Temperature Coefficient

With optional temperature compensation; valid from 10% to 100%

of full scale calibration

Flow: Maximum ± 0.015% of reading / °F up to 850 °F

[± 0.03% of reading / °C up to 454 °C]

„

Repeatability

Flow: ± 0.5% reading

Temperature: ± 1 °F [± 1 °C] (flow rate must be greater than 5 AFPS

[1,5 NMPS])

„

Turndown Ratio

Normally factory set and field adjustable from 2:1 to 100:1

within calibrated flow range; up to 1000:1 possible with factory

evaluation of application

„

Temperature Compensation

Standard: ± 30 °F [± 16 °C]

Optional: ± 100 °F [± 55 °C]

„

Agency Approvals pending

FM, FMc: Class I, Division 1, Hazardous Locations;

Groups B,C,D,E,F,G

ATEX and IECEx: Zone 1, II 2 GD Ex d IIC T4

CPA, NEPSI

„

Calibration

Performed on NIST traceable flow stands and equipment

Flow Element

„

Material of Construction

All-welded 316L stainless steel; Hastelloy-C optional

„

Operating Pressure

Metal ferrule: 1000 psig [69 bar (g)]

Teflon ferrule: 150 psig [10 bar (g)] (200 °F [93 °C] maximum)

Fixed Connection NPT: 1000 psig [69 bar (g)]

Fixed Connection Flanged: per flange rating

„

Operating Temperature (Process)

ST100, ST102 Insertion Style

All Flow Elements (– FPC, – FP and – S):

-40 °F to 350 °F [-40 °C to 177 °C]

-40 °F to 500 °F [-40 °C to 260 °C]

-40 °F to 850 °F [-40 °C to 454 °C]

ST110, ST112 Insertion Style

– FP Style Flow Element:

-40 °F to 350 °F [-40 °C to 177 °C]

-40 °F to 500 °F [-40 °C to 260 °C]

STP Series Insertion Style

All Flow Elements (– FPC, – FP and – S):

-40 °F to 257 °F [-40 °C to 125 °C]

ST100L In-line Style

– FP and – S Style Flow Element:

-40 °F to 250 °F [-40 °C to 121 °C]

„

Process Connection

Compression Fittings: Models ST100 and ST102 only

3/4” or 1” male NPT, stainless steel with adjustable Teflon ferrule or

metal ferrule; or flanged tapped and threaded for 3/4″ fitting, ANSI or

DIN flanges

Compression fittings not available with 850 °F [454 °C] temperature

versions of ST100 or ST102

Retractable Packing Glands

Low pressure 50 psig [3,5 bar (g)] or medium pressure 500 psig

[34 bar (g)] with graphite or Teflon packing material; 1 1/4″ male NPT

or ANSI or DIN flange

Teflon packing required when process media is ozone, chlorine or

bromine

Fixed Fittings / All Welded

1” male NPT or ANSI or DIN flange

Insertion Length

Field adjustable lengths:

1” to 6” [25 mm to 152 mm]

1” to 12” [25 mm to 305 mm]

1” to 21” [25 mm to 533 mm]

Technical Specification