Totalizer – Fluid Components International ST100 Series User Manual

Page 24

Advertising
background image

INSTALLATION

ST100 Series Flow Meter

20

Fluid Components International LLC

Totalizer
The Flow totalizer function accumulates the instrument total flow, very much like the odometer on an automobile. The engineering flow units

must be set in mass or volumetric units for this function to work. The total flow value is displayed directly below the real time flow value on

the instrument LCD display.

Total

Flow Value

Totalizer Setup
The ST100 totalizer function is accessed via the ST100 Configurator. Reference the ST100 Configuration supplemental manual for installation

instructions, FCI document number 06EN003403. The Configuration program is installed on a PC or Laptop computer and can be connected via

the USB or Ethernet. The USB is the primary mode of communication with the ST100.
To access the ST100 Totalizer Function:
1. Connect a USB cable between the computer with the Configuration Application installed and the USB connection on the interface board.
2. Select the ST100 Configurator icon to open the application. The application will open to the welcome screen as shown above.
3. Select the USB Connect button at the top of the screen. The Configuration application will open to the Process Data screen as shown

below.

Welcome Screen

Advertising