Fluid Components International ST100 Series User Manual

Page 29

Advertising
background image

ST100 Series Flow Meter

INSTALLATION

Fluid Components International LLC

25

Alarm Activation/Deactivation – tag AL
When an alarm condition either activates (alarm condition met) or deactivates (alarm condition no longer present), an alarm log entry will be

generated.
Following the tag field, the data items are:
• Alarm Id (1..6)
• Activated (1) or Deactivated (0)
• Alarm mode:

o Flow high: 1
o Flow low: 2
o Temperature high: 3
o Temperature low: 4
o Pressure high: 5
o Pressure low: 6

• Alarm process data value

o If Alarm mode 1 or 2, then flow (in customer units)
o If Alarm mode 3 or 4, then temperature (in customer units)
o If Alarm mode 5 or 6, then pressure (in customer units)

year month

day

time

tag

Alarm

ID

Act/

De-Act

Mode PD value fault code

fault code

fault code

2011

5

24

13:44:09

AL

1

1

2

0

0x00100000 0x00000001 0x00000000

2011

5

24

13:44:39

AL

2

1

3

86.63182 0x00100000 0x00000001 0x00000000

Log File Entry Example: Alarm 1 and 2 activated

FE Self Test: Delta-R – tag DR
When an automatic Delta-R test for an FE is executed, the results are logged with a Delta-R log entry.
Following the tag field, the data items are:
• FE Id (0..15)
• Delta-R low measured flow value (in customer units)
• Delta-R low expected flow value (in customer units)
• Delta-R medium measured flow value (in customer units)
• Delta-R medium expected flow value (in customer units)
• Delta-R high measured flow value (in customer units)
• Delta-R high expected flow value (in customer units)

year month day

time

tag

FE ID DR low m DR low e DR med m DR low e

DR hi m

DR hi e

2011

5

24

13:44:39

DR

0

60.48953

60.48953

100.48953 100.48953 150.48953 150.48953

Log File Entry Example: Automatic Delta R Calibration Check

Note: This log function is setup in the Diagnostic menu under the Scheduled Tests tab
Two other tag codes will be logged if the instrument identifies a system fault. These additional tags are FL and CF. If these fault tags are

identified in a log file, contact FCI customer service.

Advertising