2 7ma2200 series sensor wiring-si792(x) t, 2 7ma2200 series sensor wiring—si792(x) t, Installation – Hach-Lange SI792(x) E HACH GLI 3700 series SI792(x) T 7MA2200 and 8398 series User Manual User Manual

Page 29

Advertising
background image

Installation

27

3.5.3.2 7MA2200 series sensor wiring—si792(x) T
1.
Install the jumper between terminal 2 and 3 as shown in

Figure 7

.

2. Use

Table 4

to wire the sensor to the transmitter.

Figure 7 si792(x) T wiring for 7MA2200 series sensors

1 Jumper between terminal 2 and 3

2 Wiring terminals—see

Table 4

Table 4 Terminal assignments—7MA2200 series sensors

Terminal No. Assignment

Wire color

1

Receive hi

green

2

Receive lo

yellow

3

Send lo

white

4

Send hi

brown

5

Cable shield connection
Note: Do not connect to earth ground.

violet/black

7

RTD (resistive temperature device)

black/grey

8

RTD (resistive temperature device)

red/pink

Advertising