Calibration

75

8.1.4 Zero calibration—air

1. Press

CAL

, enter passcode: 1001,

ENTER

. CAL ZRO will be

displayed for 3 seconds.

2. Remove and clean the sensor.

Note: Make sure that the sensor is dry.

3. Use the

UP ARROW

and

RIGHT ARROW

to change the value until

zero is displayed as the conductivity value.
Note: If the entry of the zero point value takes longer than
6 seconds, the display will alternate between showing the
conductivity value and the temperature value.

4. Press

ENTER

to confirm.

5. The display will show the new cell constant value and the zero point.

Press ENTER to confirm.

6. Immerse the sensor in the process water.

7. The transmitter remains in HoLD mode. Press

ENTER

again.

The transmitter remains in HoLD mode for approximately
20 seconds after calibration and returns to the measure mode.

Note: Repeat the calibration when “Err” is displayed.

8.1.5 Zero calibration—calibration solution

Use a calibration solution with low conductivity for calibration.

1. Press

CAL

, enter passcode: 1001,

ENTER

. CAL ZRO will be

displayed for 3 seconds.

2. Remove and clean the sensor.

3. Immerse the sensor in the calibration solution.

4. Use the

UP ARROW

and

RIGHT ARROW

to change the value until

the conductivity value of the calibration solution is displayed.