Calibration

74

8.1.3 Calibration by comparison/grab sample

The sensor can be calibrated by measuring the process water with a
portable instrument or taking a grab sample and measuring its value in
the lab (one-point calibration). The sensor does not need to be removed
from the process during calibration. The sample temperature should
correspond to the measured process temperature for accurate results.
Note: The calibration by comparison option is not available when the
selected range and units is set to 000.0%.

1. Press

CAL

, enter passcode: 1105,

ENTER

. The display will briefly

show CAL PRD and then StorE.

2. Measure the process water with a portable instrument or collect a

grab sample for measurement in the lab. Press

ENTER

to proceed.

The cal mode indicator will flash to indicate that calibration has not
been completed. The transmitter will continue to measure and
display the current sample values.

3. Measure the grab sample with a laboratory or portable meter.
4. Press

CAL

, enter passcode: 1105,

ENTER

to access the product

calibration once more. The display will briefly show CAL PRD and
then CALC.

5. Edit the displayed value to match the value measured by the

laboratory or portable meter and press

ENTER

. The display will

show the new cell constant.

6. Press

ENTER

to end the calibration. The display will show the

measured value alternating with Hold.

Press

ENTER

to return to the measuring mode. The outputs will remain

in the hold mode for approximately 20 seconds.
Note: For comparison calibration via Foundation Fieldbus, see

section 6.2.4 on page 63

.