1 current output configuration, Section 5.2.1, Operation—4–20 ma/hart – Hach-Lange SI792(x) E HACH GLI 3700 series SI792(x) T 7MA2200 and 8398 series User Manual User Manual

Page 38

Advertising
background image

Operation—4–20 mA/HART

36

Note: During configuration the transmitter remains in the Hold mode for
safety reasons. The loop current is frozen at the value specified in the
o1.HoLD menu option. The Sensoface icon is inactive. The
configuration mode indicator is displayed (

Figure 11 on page 33

).

5.2.1

Current output configuration

Select the sensor

o1.SnSR

si792 T:

2208 SIE (7MA2200 series; default)
8398 PM
Other

si792 E:

3700 GLI (default)
Other

Enter the cell factor

1

o1.CELL

0.100–20.000 (3.00 default—si792 T; 4.70 default—si792 E)

Enter the transfer ratio

1

o1.SFC

1.00–200.00 (68.0 default—si792 T; 25.0 default—si792 E)

Select the temperature sensor

1

o1.rTD

Pt100 (default—si792 T)
Pt1000 (default—si792 E for GLI/Hach 3700 series sensors)
NTC100
NTC30

Select the range and units

o1. UnIT

0.000 mS/cm (si792 T)

Conductivity range and resolution

00.00 mS/cm
000.0 mS/cm (default)
0000 mS/cm
0.000 S/m
00.00 S/m
000.0 SAL

Salinity (SAL)

000.0%

Concentration (Conc)