4 calibration via foundation fieldbus, Operation—foundation fieldbus – Hach-Lange SI792(x) E HACH GLI 3700 series SI792(x) T 7MA2200 and 8398 series User Manual User Manual

Page 65

Advertising
background image

Operation—Foundation Fieldbus

63

6.2.4 Calibration via Foundation Fieldbus

The transmitter can be calibrated via Foundation Fieldbus using the
comparison or grab sample method.

1. Make sure the system is configured for conductivity

(PRIMARY_VALUE_TYPE = mS/cm or S/m).

2. Collect a grab sample and set CAL_SAMPLE_PRD to sample. The

conductivity value of the sample is stored. After writing, the
parameter is automatically reset to NOP (no operation).

3. Read the parameter CAL_SAMPLE_PRD_STORED_VAL. It

contains the stored value.

4. Measure the grab sample and write the lab value in

CAL_PRODUCT. The device is now calibrated.
CAL_SAMPLE_PRD_STORED_VAL is reset to zero.