Operation—4–20 mA/HART

38

Note: During configuration the instrument/transmitter remains on Hold.
Depending on the configuration the loop current is frozen at its last
value or at a preset fixed value.

Press

ENTER

to access a menu item. Use the

ARROW KEYS

to edit

values. Press

ENTER

to save the settings. If a value is outside of the

acceptable range, “Err” will be displayed and the value will not be
accepted. To exit the menu and return to the measurement mode, press

CONF

and

ENTER

.

Example: Set the output start point for the 4 mA signal to be 10.0 mS
and the output end point to be 100 mS for the 1.20 mA signal.
1. Press

CONF

, enter passcode:

1200

. The display will show Conf and

then out.1MNU.

2. Press

ENTER

to access the output setup menu. The display will

show

CELL

.

3. Press

ENTER

,

ENTER

,

ENTER

to reach the 4 mA submenu. The

display will show o1.4mA.

4. Use the

UP ARROW

and

RIGHT ARROW

to edit the value to read

10.0 mS value. Press

ENTER

to save the value. The display will

show o1.20mA.

5. Use the

UP ARROW

and

RIGHT ARROW

to edit the value to read

100.0 mS value. Press

ENTER

to save the value. The display will

show o1.FtME.

Specify the value to output during HOLD periods

o1.FIX

21.0 mA (default)

During HOLD the entered value is
maintained at the output (refer to

section 5.2.1.3

for more information)

1

If “Other” sensor is selected, o1.CELL, o1.SFC and o1.rTD are available

2

If the process variables SAL, % (Conc) or USP are selected the following steps

(Lin/Log) are omitted.

5.2.1

Current output configuration (continued)