Page 52: Operation—foundation fieldbus

Operation—Foundation Fieldbus
50
6.2.2 Standard analog input block (AI)
Three Analog Input Function Blocks provide for cyclic transmission of
measured values (currently measured value with status, alarm limits,
freely selectable process parameter).
6.2.2.1 Operating modes
Use the MODE_BLK parameter to set the following operating modes:
•
OOS—out of service. If not write-protected, access to all
parameters is allowed.
•
MAN—manual
•
Auto—online, normal state
6.2.2.2 Set the parameter and units
Use CHANNEL to set the measured parameter and units (
). The
corresponding measurement unit is selected in the UNITS
subparameter of XD_SCALE (
Table 9 on page 52
).
Table 8 Measurement parameters and units
Channel
Parameter
Unit
Unit value
1
Conductivity
µS/cm
1586
mS/cm
1302
S/cm
1299
2
Concentration
% (percent)
1342
3
Temperature
°C
1001
°F
1002
4
Salinity
‰ (parts per thousand) 2003
5
Resistance
M
Ω/cm
1587
6
Cell constant
cm
-1
2004