Operation—Profibus PA

66

7.1.2

Configuration menu

Select the range and units

0.000 mS
00.00 mS
000.0 mS (default)
0000 mS

000.0%

-01- NaCl (default)
-02- HCl
-03- NaOH
-04- H

2

SO

4

(0 – 35% by weight)

-05- HNO

3

-06- H

2

SO

4

(95 – 99% by weight)

000.0 SAL

Select temperature unit

°C (default)
°F

Select temperature sensor

100 PT (default)
1000 PT
30 NTC
100 NTC
busEXT (external temp during measurement; manual temp during
calibration, °C)

Select temperature compensation

1

tc

OFF

Temperature compensation turned off

LIN (default)

Linear temperature compensation with entry
of temperature coefficient (00.00–19.99%/
K). Reference temperature = 25 °C.
Enter temperature coefficient
02.00%/K (default) (xx.xx%/K)

nLF

Temperature compensation for naturals
waters to EN 27888