2 configuration menu, Section 7.1.2, Operation—profibus pa – Hach-Lange SI792(x) E HACH GLI 3700 series SI792(x) T 7MA2200 and 8398 series User Manual User Manual
Page 68
Advertising

Operation—Profibus PA
66
7.1.2
Configuration menu
Select the range and units
0.000 mS
00.00 mS
000.0 mS (default)
0000 mS
000.0%
-01- NaCl (default)
-02- HCl
-03- NaOH
-04- H
2
SO
4
(0 – 35% by weight)
-05- HNO
3
-06- H
2
SO
4
(95 – 99% by weight)
000.0 SAL
Select temperature unit
°C (default)
°F
Select temperature sensor
100 PT (default)
1000 PT
30 NTC
100 NTC
busEXT (external temp during measurement; manual temp during
calibration, °C)
Select temperature compensation
1
tc
OFF
Temperature compensation turned off
LIN (default)
Linear temperature compensation with entry
of temperature coefficient (00.00–19.99%/
K). Reference temperature = 25 °C.
Enter temperature coefficient
02.00%/K (default) (xx.xx%/K)
nLF
Temperature compensation for naturals
waters to EN 27888