1 enter cell constant, Calibration – Hach-Lange SI792(x) E HACH GLI 3700 series SI792(x) T 7MA2200 and 8398 series User Manual User Manual

Page 74

Advertising
background image

Calibration

72

o1.HoLD menu option. The Sensoface icon is inactive. The calibration
mode indicator is displayed (

Figure 11 on page 33

).

To exit the calibration mode at any time, press

CAL

and then

ENTER

.

The output current will be held for 20 seconds and the measured value
will be displayed.

8.1.1

Enter cell constant

1. Press

CAL

, enter passcode: 1100,

ENTER

. CAL CELL will be

displayed for 3 seconds.

2. Use the

UP ARROW

and

RIGHT ARROW

to enter the value of the

cell constant of the connected sensor. The conductivity value will be
displayed also.
Note: A cell constant change will change the conductivity value.

Note: If the entry of the cell constant takes longer than 6 seconds,
the display will alternate between showing the conductivity and
temperature value.

3. Press

ENTER

to save the value.

4. The transmitter displays the cell constant value and CELL. Press

ENTER.

5. The transmitter remains in HoLD mode. Press

ENTER

again.

The transmitter remains in HoLD mode for approximately
20 seconds after calibration and returns to the measure mode.

Note: Repeat the calibration when “Err” is displayed.