3 8398 series sensor wiring-si792(x) t, 3 8398 series sensor wiring—si792(x) t, Installation – Hach-Lange SI792(x) E HACH GLI 3700 series SI792(x) T 7MA2200 and 8398 series User Manual User Manual
Page 30
Advertising

Installation
28
3.5.3.3 8398 series sensor wiring—si792(x) T
1. Install the jumper between terminal 2 and 3 as shown in
Figure 8
.
2. Use
to wire the sensor to the transmitter.
Figure 8 si792(x) T wiring for 8398 series sensors
1 Jumper between terminal 2 and 3
Table 5 Terminal assignments—8398 series sensor
Terminal No. Assignment
Wire color
1
Receive hi
brown
2
Receive lo
black
3
Send lo
shield of brown wire
4
Send hi
shield of white wire
5
Cable shield connection
Note: Do not connect to earth ground.
white
7
RTD (resistive temperature device)
green
8
RTD (resistive temperature device)
yellow
Advertising