Section 6 troubleshooting, 1 error messages, 2 warnings – Hach-Lange RTC103 N-Module User Manual

Page 55: 3 wear parts, 1 error messages 6.2 warnings 6.3 wear parts

Advertising
background image

55

Section 6

Troubleshooting

6.1

Error messages

Possible RTC errors are displayed by the sc controller.

6.2

Warnings

Possible RTC sensor warnings are displayed by the sc controller.

6.3

Wear parts

Displayed errors

Definition

Resolution

RTC MISSING

No communication between RTC and RTC
communication card

Supply RTC with voltage
Test connection cable
Reset the sc1000 and the RTC (switch so it is
completely voltage free and switch back on)

RTC CRC

Interrupted communication between RTC and
RTC communication card

Make sure +/- connections of the connector
cable between RTC and RTC communication
card in the sc1000 are installed correctly.
Change, if necessary.

CHECK KONFIG

The sensor selection of the RTC was deleted
by removal or selection of a new sc1000
participant.

From

MAIN MENU > RTC MODULES /

PROGNOSYS > RTC MODULES > RTC >
CONFIGURE > SELECT SENSOR

, select the

correct sensor for the RTC again and confirm.

RTC FAILURE

Brief general read/write error on the CF card,
mostly caused by a brief interruption to the
power supply.

Acknowledge error. If this message is shown
frequently, eliminate the cause of the power
disruptions. If necessary, inform the service
team of the manufacturer (

Section 8

).

Displayed warnings

Definition

Resolution

MODBUS ADDRESS

The RTC menu SET DEFAULTS was opened.
This deleted the Modbus address of the RTC in
the sc1000.

MAIN MENU > RTC MODULES / PROGNOSYS >
RTC MODULES > RTC > CONFIGURE > MODBUS
> ADDRESS

: Access this menu and set the

correct MODBUS address.

PROBE SERVICE

A configured sensor is in service status.

The sensor must exit service status.

Component

Quantity

Service life

CF card, type for RTC module

1

2 years

Battery, type CR2032 Panasonic or Sanyo

1

5 years

Advertising