11.

OPERATION

11.1.

ZEROING

You can press the [Zero] key at any time to set the zero point from
which all other weighing and counting is measured. This will usually be
necessary when the platform is empty. When the zero point is obtained
the display will show the zero indicator.

The indicator has an automatic re-zeroing function to account for minor
drifting or accumulation of material on a connected platform. However
you may need to press [Zero] to re-zero the indicator if small amount of
weight is still shown when the platform is empty.

11.2.

TARING

11.2.1

MANUAL TARE

Zero the indicator by pressing [Zero]. The zero indicator will be on.
Place a container on the pan and its weight will be displayed.

Press [Tare] when the reading is stable. The weight that was displayed
is stored as the tare value and it is subtracted from the display, leaving
zero on the display. The stable and Net indicator will be on.

As a product is added only the weight of the product will be shown. The
indicator could be tared a second time if another type of product was to
be added to the first one. Again only the weight that is added after
taring will be displayed.