24 © Adam Equipment Company 2011

NOTE:

1.

If the recalled limit is for parts counting, the display will show the last sample
value used, ready for a new sample to be counted.

2.

If the recalled unit is a percent weighing limit, the display will show the last
sample value used, ready for a new sample to be weighed.

3.

If the memory location was empty the indicator will return to weighing.

11.7.

PERCENT WEIGHING

The indicator can be set to perform percent weighing. See Section 13.1.

The indicator will use a mass on the platform as the 100% reference weight. If the

platform is empty (or the indicator is tared) then the user can input a reference

weight using the keypad.

If using a reference weight (or object) as your 100% reference, add the weight
to the to the platform.

Press [Func]. The first option is “FUnC 1”, press the [Func] key 3 more times to
display “FUnC 4”.

Press the [Tare] key. “F4 PCt” will be displayed.