SERVICE PARAMETERS

16.1.

ACCESS TO PARAMETERS

APPROVED INDICATORS

Access to the indicator parameters and calibration is controlled in all approved

indicators either by limiting access to be after the Calibration Jumper is put on the

PCB, location J1, pins 1 & 2. In this case the display will show the passcode request

screen, “ P - - - - “ . To continue enter a passcode as described below.

Or if the Calibration and Parameters have been enabled (see 15.2.10) the user must

enter the correct password to have access. See Section 6.0.

Entering passcode 0000 will allow calibration as shown in 15.1, entering 1000 will

allow access to a limited set of parameters as described in section 6.2 and entering

the passcode 2006 will allow access to all parameters as shown in section 15.2.

NON-APPROVED INDICATORS

Non EC Approved indicators will allow entry to the parameters if the Tare key is

pressed during the power on cycle. The passwords work as above.

15.1 USING “0000” TO ENTER THE CALIBRATION PARAMETER

“Pn”

When “Pn” is displayed.

Enter “0 0 0 0” and press [Tare]

“UnLoAd ”

Empty the platform by removing the load, if

there is any and press [Tare]

“LoAd” “6” “KiLoS”

Load the requested calibration weight and press

[Tare]