Nova-Tech GKa Indicator User Manual

Page 59

Advertising
background image

57 © Adam Equipment Company 2011

16.2.10

F10 –CALIBRATION COUNT (GK-M ONLY)

This parameter allows the calibration and parameter counting function to be active.

To enter this parameter, press the [Tare] key when “F10 Cn” is shown.

The display will show if the Calibration Counting Mode is set to on or oFF. If On the

Calibration count and Parameter count will be seen at power on as described in

section 6.0. If set to Off the only method that can be used for access to the

calibration or parameters is to place the jumper on pins 1-2 of J1, See section 15.2.

Press the [Func] key to change the value.

Press [Tare] to accept the displayed value.

Press [Zero] to return to weighing.

16.2.11

F10 –AUTO ZERO RANGE (NOT AVAILABLE ON GK-M)

This parameter set the range the autozero is active within.

To enter this parameter, press the [Tare] key when “F10 Az” is shown.

The display will show the current setting, 0.5d, 1d, 2d, or 5d.

Press the [Func] key to change the value.

Press [Tare] to accept the displayed value.

Press [Zero] to return to weighing.