52 © Adam Equipment Company 2011

16.2.1

F1 -CALIBRATION

To enter the calibration parameter, press the [Tare] key when “F1” is displayed. The

indicator will be calibrated using 2 masses of approximately 1/3Maximum and

Maximum. If the indicator has been calibrated once the values will be stored. If this

is the first time the indictor is calibrated the user must enter the values for the

calibration masses.

It is necessary to set the decimal point location and the capacity before calibration is

possible.

The display will instruct you to remove any weight from the platform, “UnloAd”.

Press [Tare].

The display will tell you to add the first weight to the platform: “Ld 1” “ 10 kg” If

necessary change the value shown to match the weight to be used. Press [CE] to

clear the old value and then enter the new value. All values entered are in whole

numbers only.

Add the weight shown, wait for stability then press the [Tare] key.

The display will tell you to add the second weight to the platform: “Ld 2 “ “ 30 kg”

Add the weight shown, wait for stability then press the [Tare] key.

The display will show “SPAn” “PASS” if the calibration is OK.

Remove the weight.

For the approved indicator the display will then show “JP” “On” indicating the jumper

is still in place if the jumper within the indicator was used to access the parameters..