© Adam Equipment Company 2011

5

2.

SPECIFICATIONS

INPUT SECTION

Load Cells

Up to 4 , 350 ohm load cells
Minimum 87 ohms, maximum 1120 ohms

Connection

6 wires
2 excitation, 2 sense, 2 signal

Excitation

5Vdc

Sensitivity

0.15uv/d (GK-M, 1.5uv/e)

Linearity

0.01% FS

Zero Range

0- 10mv

Signal range

0-40mv

ADC Sensitivity

Approximately 0.02 µv/ADC count

DIGITAL SECTION
Maximum Range

Typically 1kg – 30000kg

Divisions

Up to 30,000, (GK –M, 3000 or less)

Weigh units

kg / g for Europe, Asia, Africa
kg / g / lb / oz / lb:oz for USA
kg only for GK-M series

Stabilisation Time

2 Seconds typical

Operating
Temperature

-10°C - 40°C
14°F - 104°F

Power supply

230 VAC 50/60 Hz
12V 800ma adaptor for USA verisons

Battery

Internal rechargeable battery

Calibration

Automatic External

Display

6 digits LCD digital displays
with capacity tracker and symbols for units

Indicator Housing

ABS Plastic

Overall
Dimensions
(wxdxh)

260 x 170 x 115mm
10.2” x 6.7” x 4.5”

Net Weight

2 kg / 1 lb

Applications

Weighing and check weighing

Functions

Weighing, Check Weighing, Parts counting, check-counting, , Animal
Weighing, Accumulating memory,

Interface

RS-232 bi-directional interface
English, German, French, Spanish selectable text

For approved indicators the input specifications is limited to 1.5 µv per division and

the number of divisions is limited to 3000d. Kilograms only.