Nova-Tech pH 110 User Manual

Page 20

Advertising
background image

Instruction Manual

CyberScan pH 11 / 110

16

4.5 Temperature

Calibration

The temperature sensor is factory calibrated. Calibrate
the temperature probe only if you suspect temperature
errors may have occurred over a long period of time or if
you have a replacement temperature probe. This
procedure offers offset adjustment of probe to ensure
more accurate temperature measurement.

1. Make sure the ATC probe (or temperature

connector of the electrode) is connected to the
phono-socket (see Figure 6).

2. Switch the meter on. The ATC annunciator will

appear at the right-hand side of the LCD. Press
the MODE key to select pH mode.

3. Press the CAL/MEAS key to enter pH

calibration mode. The CAL indicator will appear
above the primary display.

4. While in pH calibration mode, press the MODE

key to enter temperature calibration mode. The
primary display shows the measured
temperature reading according to the last set
offset

1

and the secondary display shows the

factory calibrated temperature measurement.

5. Dip the ATC probe (or electrode) into a solution

of known temperature (i.e. a temperature bath).
Allow time for the temperature probe to
stabilise.

6. Scroll with the MI/Ÿ or MR/ź keys to set the

correct temperature value (i.e. the temperature
of the temperature bath). You can adjust the
reading in increments of 0.1°C.

7. Once you have selected the correct

temperature, press the HOLD/ENTER key. The
meter automatically returns to pH measurement
mode. See Figure 14.

NOTES:

You can offset the temperature reading up to ±5°C from
original reading.

To exit this program without confirming the temperature
calibration value, DO NOT press ENTER. Press CAL/MEAS instead

.

1

It displays the value being measured currently, offset by the last calibration.

MEAS

pH

ATC

°C

7.16

22.3

MEAS

pH

ATC

°C

7.16

22.0

CAL

pH

pH

7.16

7.00

CAL

ATC

°C

22.3

22.3

ATC

°C

22.0

22.3

CAL

READY

MEAS

CAL

MODE

MI

MR

ENTER

HOLD

Figure 14: Temperature

calibration

http://www.novatech-usa.com/Products/Laboratory-Equipment-Supply-Products Tel: (281) 359-8538 Toll Free:(866) 433-6682

Advertising
This manual is related to the following products: