Instruction Manual

CyberScan pH 11 / 110

42

9.6

P5.0 : Reset to Factory Default

This mode lets you reset all parameters to factory
default settings. There are two levels of reset:

x Calibration Reset: Reset only the calibration

values. This clears all previous calibrated
values and reset to factory default. See page 68
and 69 for factory default settings.

x User Reset: Clears all data including

calibration, memory and other customised
setup functions and reset it to factory default.
See page 68 and 69 for factory default settings.

9.6.1 Calibration

Reset

From measurement mode:

1.

Press SETUP key to enter Set Up mode.

2.

Press the MI/Ÿ or MR/ź keys to scroll
through subgroups until you view parameter
P5.0.

3.

Press the HOLD/ENTER key.

4.

Press the MI/Ÿ or MR/ź keys to toggle
between NO and YES.

x ‘NO’ retains current settings

x ‘YES’ erase calibration settings.

5.

Press the HOLD/ENTER key to confirm
selection.

x If ‘NO’ is selected, it will retains current

settings and proceed to next program
P5.2 – User Reset.

x If ‘YES’ is selected, all LCD segments will

lit up for one second, erase calibration
values, and the meter switches to the
measurement mode.

6.

Otherwise press CAL/MEAS key to return to
measurement mode without resetting to factory default.

See Figure 38.

yes

SETUP

SETUP

SETUP

SETUP

P 5.0

P 5.1

CAL

CAL

CAL

ENTER

HOLD

ENTER

HOLD

MI

MR

Select either "YES" or "nO",
press ENTER key to confirm.

Figure 38: P5.1 - Calibration

reset