12 error messages – Nova-Tech pH 110 User Manual

Page 64

Advertising
background image

Instruction Manual

CyberScan pH 11 / 110

60

12 ERROR

MESSAGES

The following table provides a guideline to enable diagnosis of possible problems
indicated by the messages generated by the meter. The table also provides possible
solutions to the problems encountered.

ERROR MESSAGE

INDICATES

POSSIBLE CAUSE

CORRECTIVE
ACTION

Err. Annunciator

Wrong keypad input

Wrong input in
selected mode.

Release key.
Select valid
operations
depending on
mode.

Secondary Display
Toggles Continuously
between Calibration
Buffer Values,
Electrode icon blinks &
Err annunciator lights
up

Calibration

Incorrect buffer used
or contaminated buffer
solution

Check if the right
buffer was
selected or use
fresh buffer
solution.

Battery icon lights up

Low Battery

Battery power is low

Replace batteries
with fresh ones as
soon as possible.

http://www.novatech-usa.com/Products/Laboratory-Equipment-Supply-Products Tel: (281) 359-8538 Toll Free:(866) 433-6682

Advertising
This manual is related to the following products: