9.4

P3.0: View Electrode Data

Program 3 has two “view only” options that let you check the electrode parameters
for diagnostic purposes.

x From pH measurement mode, you can view the electrode’s offset and slope

values.

x From mV measurement mode, you can view the electrode’s relative mV offset

value. (pH 110 only)

9.4.1 View pH Electrode Offset and Slope

From pH measurement mode:

1.

Press the SETUP key to enter Set Up mode.

2. Press the MI/Ÿ or MR/ź keys to scroll through

subgroups until you view parameter P3.0.

3.

Press the HOLD/ENTER key.

4. The display shows the electrode offset value. It

is the mV offset at pH 7.00. If you have not
calibrated at any buffer, the primary display
shows ‘---‘.

5. Press the HOLD/ENTER key to proceed to

electrode slope display.

6. The display shows electrode slope in

percentage. Slope displayed will be the current
slope of the slope zone to which a measurement
is made or calibration is done.

7. At any point, you can press the CAL/MEAS key

to return to measurement mode.

See Figure 35.

Figure 35: P3.1/3.2 - View pH

electrode offset and slope

mV

pH

SETUP

5.3

p3.1

SETUP

p 3.0

ele

%

pH

SETUP

p 3.2

100.0

ENTER

HOLD

ENTER

HOLD