P 5.2 – Nova-Tech pH 110 User Manual

Page 47

Advertising
background image

Instruction Manual

CyberScan pH 11 / 110

43

NOTE:

Calibration reset is independent for each measurement parameter.
That is:
x From pH mode, calibration reset will only erase pH calibration

values; and

x From mV mode, calibration reset will only erase mV calibration

values.

9.6.2 User

Reset

This program clears all data including calibration,
memory and other customised setup functions and reset
it to factory default.

You can skip program P5.1 (Calibration Reset) and
proceed straight to Program P5.2:

1. From Program P5.0 - rSt, press HOLD/ENTER

key three times to P5.2.

2. Press the MI/Ÿ or MR/ź keys to scroll through

subgroups until you view parameter P5.0.

x

‘NO’ retains current settings

x

‘YES’ resets to factory default settings.

3.

Press the ENTER key to confirm selection.

x If ‘NO’ is selected, it will retains current

settings and goes back to program 5.0.

x If ‘YES’ is selected, all LCD segments will lit

up for one second, resets to factory default
settings, and the meter will be in the
measurement mode.

4. Otherwise press CAL/MEAS key to return to

measurement mode without resetting to factory
default.

yes

SETUP

SETUP

SETUP

P 5.2

ENTER

HOLD

MI

MR

Select either "YES" or "nO",
press ENTER key to confirm.

Figure 39: P5.2 - User reset