P 1.2 – Nova-Tech pH 110 User Manual

Page 39

Advertising
background image

Instruction Manual

CyberScan pH 11 / 110

35

9.2.2 P1.2: Select Calibration Buffer Options

The types of pH calibration buffer options available are:

Meter

Meter Display

pH Calibration Buffer Options

pH 11

USA

nSt

USA buffers --- pH 1.68, 4.01, 7.00, 10.01, and 12.45.

NIST buffers --- pH 1.68, 4.01, 6.86, 9.18, and 12.45.

pH 110

USA

nSt

dn

PUr

USA buffers --- pH 1.68, 4.01, 7.00, 10.01, and 12.45.

NIST buffers --- pH 1.68, 4.01, 6.86, 9.18, and 12.45.

DIN buffers --- pH 1.09, 3.06, 4.65, 6.79, 9.23, and 12.75.

PWB ( Low Ionic): pH 4.10 and 6.97

From measurement mode

Press SETUP key to enter Set Up mode.

1. Press the MI/Ÿ or MR/ź keys to

scroll through subgroups until you
view parameter P1.0.

2. Press the HOLD/ENTER key two

times to select parameter 1.2.

3. Press the MI/Ÿ or MR/ź keys to

select the buffer set you require:

4. Press the HOLD/ENTER key to

confirm selection and to return to the
subgroup menu. Press CAL/MEAS
key to return to measurement mode.

See Figure 31.

This screen is
available only
in the pH 110
meter

This screen is
available only
in the pH 110
meter

SETUP

SETUP

usa

p 1.2

p 1.2

SETUP

p 1.2

SETUP

p 1.2

MR

MI

MR

MI

MR

MI

Figure 31: P1.2 - Select pH calibration

buffer standards

http://www.novatech-usa.com/Products/Laboratory-Equipment-Supply-Products Tel: (281) 359-8538 Toll Free:(866) 433-6682