CyberScan pH 11 / 110

25

8.2.1 Print Current Data Manually

To manually print any data that is currently being
measured:

1. From measurement mode, press HOLD key to

freeze reading. See Figure 24.

2.

Print PRINT (

) key.

3. The “Prn – CUr” annunciator will appear and

flash for about 0.5 seconds. The data is
transmitted to the printer or computer.

4.

The meter returns to measurement mode.

5.

Repeat steps 1 to 3 for next measurement.

8.2.2 Print Data on Timed Interval

This feature is useful to automate printing and
downloading ‘live’ measurement values into computer.
The time interval setting can range from 5 to 30
seconds.

1. From pH or mV measurement mode, press

PRINT key.

2.

Select “Dat/Cur” screen and press
HOLD/ENTER key.

3. The “Sec 5” screen appears. The default value

is 5 seconds time interval. See Figure 25.

4.

Press MI/Ÿ or MR/ź key to set a value suitable for your application.

5.

Press HOLD/ENTER key.

6.

The meter goes into measurement mode with MEM annunciator flashing.

7. At a pre-set time interval, the screen flashes “Prn/CUr” momentarily (printing

data) and switches back to measurement mode with MEM annunciator
flashing.

8.

To stop printing, press PRINT key. The MEM annunciator disappears and the
meter is on measurement mode.

pH

pH

MEAS

MEAS

READY

READY

23.4

23.4

6.55

6.55

ENTER

HOLD

°C

°C

pH

MEAS

READY

HOLD

Flash for 0.5 seconds

23.4

6.55

°C

Figure 24: Print current data

manually